Research of a SPR Immunoassay Method for Okadaic Acid Contamination of Red Tide Toxins

Xiangdong ZHOU,Huibin CHEN,Shuyue ZHAN,Xiaoping WANG,Huichao OU,Zhaofeng LUO
DOI: https://doi.org/10.3969/j.issn.1004-1699.2011.12.026
2011-01-01
Abstract:A novel assay method for okadaic acid (OA) of red tide toxins has been exploited, which employed surface plasma resonance(SPR) technique and immunoassay method. Spreeta™ sensor was adopted in the SPR flow injection analysis(FIA) system. The reusable immunosensor chip was functionalized with okadaic acid-bovine serum albumin (OA-BSA) using covalent-coupling. Acceptable repeatability was achieved with relative standard deviation (RSD) 1.51% (n=12). Calibration curve obtained for OA has detection limit of 19.4 ng/mL, half maximal inhibitory concentration (IC 50) of 177 ng/mL and quantitative range of 40 ng/mL ∼ 640 ng/mL. The method owns a desirable repeatability. Using self-designed reusable biochip, it has an advantage of cost-effectiveness. The method has a good prospect of on-site determination of OA.
What problem does this paper attempt to address?