Detection of endogenous cortisol in human hair by high performance liquid chromatography tandem mass spectrometry

Qiaozhen Xie,Wei Gao,Jifeng Li,Ting Qiao,Huihua Deng,Zuhong Lu
DOI: https://doi.org/10.3969/j.issn.1001-0505.2011.03.029
2011-01-01
Abstract:A new method for detection of endogenous cortisol in human hair was developed based on high performance liquid chromatography tandem mass spectrometry (HPLC/MS/MS), and the results detected was compared with that determined by high performance liquid chromatography with fluorescence detector (HPLC/FLU). Hair sample (50 mg) was incubated in methanol after being washed and pulverized, and then the extraction was purified by C18 solid phase extraction column. Cortisol identification was performed by electro-spray ionization (ESI) in negative mode using multiple reaction monitoring (MRM) and the quantification was performed using external standard method. The limit of quantification is 2 pg/mg, and the linear range is 2 to 100 pg/mg. The recovery of cortisol spiked at 50 pg/mg is (99.4±9.1)%, intra-day precision is 7.6%, inter-day precision is 8.3%. Because of the differences in hair pretreatment method, cortisol extraction method, matrix effect and quantitative method, cortisol concentration detected by HPLC/MS/MS is significantly lower than that determined by HPLC/FLU. However, Pearson correlation analysis showed that there is a significant linear correlation (r=0.749, p<0.01) between hair cortisol contents analyzed by these two methods.