Construction and evaluation of continuous two-dimensional open-tubular ion exchange/reversed phase monolithic column capillary electrochromatography system

ZHU Ya-Xian,ZHANG Ling-Yi,Wen Lei,ZHANG Wei-Bing
2011-01-01
Abstract:A continuous two-dimensional open-tubular ion exchange/reversed phase monolithic column capillary electrochromatography system was developed. Reversed phase monolithic column was firstly synthesized in the end part of the capillary, then N-[3-(trimethoxysilyl)-propyl] -ethlenediamine was surface coated to the other part of the capillary. The separation of seven organic acids verified the effect on the separation of ion exchange in the open-tubular capillary. Rhizoma gastrodiae was separated to evaluate the system, and 34 peaks were recognized. This system is designed for complex samples with highly hydrophobicity components and highly hydrophilicity components simultaneous. This new column system combining advantages of multidimensional separation and adjusted stationary phase selectivity is suitable for the analysis method development of complicated samples.
What problem does this paper attempt to address?