Design and Development of Blood Pressure Measurement Base on DSP

Xi-Tian PI,Shuang-Shuang LI,Hong-Ying LIU
DOI: https://doi.org/10.3969/j.issn.0258-8021.2011.03.018
2011-01-01
Abstract:This paper introduces electronic sphygmomanometer' software and hardware implementation based on DSP. It makes full use of the internal and external resources of TMS320VC5402 chip and possesses the power management design. This system has such characteristics: low power, portable, simple operation procedures by using touch screen, remote monitoring of blood pressure information through USB and the PC. Oscillometric method was used to get the original detecting data which are corrected by gold standard and temperature compensation algorithm. 38 normal subjects participated in the verifying experiments. Blood pressures of them were measured with our system and the traditional sphygmomanometer in the same period. Statistical analysis was used to find the correlation between the two. Results showed that the measurement accuracy reaches 86.9% and has good correlation with traditional sphygmomanometers.
What problem does this paper attempt to address?