Electrogenerated Chemiluminescence Sensor for the Determination of Propranolol Hydrochloride

Honglan Qi,Chen Wang,Rui Zou,Libo Li
DOI: https://doi.org/10.1039/c0ay00591f
IF: 3.1
2011-01-01
Analytical Methods
Abstract:An electrogenerated chemiluminescence (ECL) sensor for the determination of propranolol hydrochloride was fabricated by employing tris(2,2'-bipyridyl)ruthenium(H) (Ru(bpy)(3)(2+)) as an ECL signal producer and graphene as a modified material. The ECL sensor was fabricated by adsorbing Ru(bpy)(3)(2+) onto a mixture of graphene and Nafion. The introduction of conductive graphene into Nafion was found to greatly enhance the ECL intensity of Ru(bpy)(3)(2+) and the ECL intensity of Ru(bpy)(3)(2+) decreased in the presence of propranolol hydrochloride. Based on this phenomenon, a sensitive and stable ECL method was developed for the determination of propranolol hydrochloride by employing a Ru(bpy)3(2+)-graphene-Nafion modified electrode. Under the optimized conditions, the ECL intensity was linear with the concentration of propranolol hydrochloride over a range of 1.0 x 10(-10) to 5.0 x 10(-9) mol L-1 with a detection limit of 3 x 10(-11) mol L-1. The ECL sensor exhibited long-term stability with a relative standard deviation of 4.1% for sixteen continuous determinations of 2.0 x 10(-9) mol L-1 propranolol hydrochloride. The developed method has been successfully applied to the determination of propranolol hydrochloride in pharmaceutical samples.
What problem does this paper attempt to address?