Analysis of Superimposed Ultrasonic Guided Waves in Long Bones by the Joint Approximate Diagonalization of Eigen-Matrices Algorithm

Xiaojun Song,Dean Ta,Weiqi Wang
DOI: https://doi.org/10.1016/j.ultrasmedbio.2011.06.028
IF: 3.694
2011-01-01
Ultrasound in Medicine & Biology
Abstract:The parameters of ultrasonic guided waves (GWs) are very sensitive to mechanical and structural changes in long cortical bones. However, it is a challenge to obtain the group velocity and other parameters of GWs because of the presence of mixed multiple modes. This paper proposes a blind identification algorithm using the joint approximate diagonalization of eigen-matrices (JADE) and applies it to the separation of superimposed GWs in long bones. For the simulation case, the velocity of the single mode was calculated after separation. A strong agreement was obtained between the estimated velocity and the theoretical expectation. For the experiments in bovine long bones, by using the calculated velocity and a theoretical model, the cortical thickness (CTh) was obtained. For comparison with the JADE approach, an adaptive Gaussian chirplet time-frequency (ACGTF) method was also used to estimate the CTh. The results showed that the mean error of the CTh acquired by the JADE approach was 4.3%, which was smaller than that of the ACGTF method (13.6%). This suggested that the JADE algorithm may be used to separate the superimposed GWs and that the JADE algorithm could potentially be used to evaluate long bones.
What problem does this paper attempt to address?