Quickly Quantitative Detection of Acrylonitrile′s Biocatalysis Products by Gas Chromatography
WANG Tie-gang,LUO Hui,ZHU Yan-qin,YU Hui-min,YANG Hui-fen,SHEN Zhong-yao
DOI: https://doi.org/10.3969/j.issn.1000-0720.2007.01.013
2007-01-01
Abstract:Products of acrylamide from acrylonitrile with microbial enzymes as the biocatalyst has recently become po-(pular) in industry.During the biotransformation process,substrate acrylonitrile,product acrylamide and by-product acrylic acid will be simultaneously present in the reaction solution.In this study,a capillary column PEG-20M(30 m×0.25 mm i.d.,2 μm) was selected to quantitatively measure the concentrations of acrylonitrile,acrylic acid and acrylamide,with acetamide as the inner standard.The optimal temperatures of the SPL sample receiver,the FID detector and the column were determined to be 260,260 and 190 ℃,respectively,with nitrogen as the carrier at the speed of 25 cm/min.The injection volume of the sample was set at 0.4 μL with the shunting ratio of 50∶1.Final results showed that the three targets can be completely separated within 4 min,with high reproducibility and linearity.The reoveries of acrylonitrile,acrylic acid and acrylamide reached as high as 98.5%,102.1% and 105.0%,with the relative standard deviation of 9.1%,5.3% and 1.7%,respectively.These results show this method is simple,accurate and has a very promising prospect in the high-throughput quantitative analyses of acrylonitrile,acrylamide and acrylic acid in both lab and industry.