Rapid quantitative determination of EPA and DHA in fish oil by ATR-FTIR with hybrid chemometrics

Di Wu.,Fang Cao,Xiaojing Chen,Yong He,Da-Wen Sun
2011-01-01
Abstract:A rapid and nondestructive procedure based on Attenuate Total Reflectance Fourier Transform Infrared Spectroscopy (ATR-FTIR) models was developed to determine the content of EPA and DHA in fish oil to replace the time consuming gas chromatography (GC) method. A total of 35 samples were scanned in the ATR mode by Nicolet 670FT-IR spectrometer and the reference values for EPA and DHA was measured by GC. Calibration equations were developed using partial least squares (PLS), least-squares support vector machine (LS-SVM) and multiple linear regression (MLR). High averaged RPD values of 6.037 and 10.597 from PLS and LS-SVM were obtained by full range spectra for DHA and EPA analysis respectively. In order to eliminate useless variables and reduce calculation time, successive projections algorithm (SPA) was used to select effective variables. The averaged RPD values of 4.657 and 5.476 from three calibration algorithms were obtained based on SPA selected variables for DHA and EPA respectively. Less than 0.5% of the full range spectral variables were selected respectively for EPA and DHA analysis, which shows that SPA was a useful tool to select the effective variables. The results reported in this article demonstrate that ATR-FTIR measurements can serve as a rapid method to determine EPA and DHA in fish oils.