Method Development For Analysis Of Short- And Long-Chain Perfluorinated Acids In Solid Matrices

Fei Li,Chaojie Zhang,Yan Qu,Jing Chen,Xiang Hu,Qi Zhou
DOI: https://doi.org/10.1080/03067310903191721
2011-01-01
International Journal of Environmental Analytical Chemistry
Abstract:In order to help to elucidate the transport and fate of perfluorinated acids (PFAs) in the environment, a reliable and sensitive analytical method has been developed in present study for determination of short-and long-chain PFAs in various solid matrices. The method consisted of solvent extraction of PFAs from solid matrices using sonication, solid phase extraction (SPE) using weak anion exchange (WAX) cartridges, clean-up of SPE eluent with dispersive carbon sorbent and quantitation by high performance liquid chromatography-negative electrospray-tandem mass spectrometry (HPLC-negative ESI-MS/MS). The method detection limits (MDL) and quantitation limits (MQL), which were analyte- and sample-dependent, ranged from 0.02 to 0.06 ng g(-1) and 0.10 to 0.90 ng g(-1), respectively. The recoveries of all PFAs were generally good enough for quantitative analysis of these chemicals (57-115%), especially for short-chain (<C8, 80-115%) PFAs excluded in previous studies because methods were not available. The precisions of this method, represented by the percent relative standard deviation (RSD) of spiked measurements, were in a range of 1-19%. In addition, matrix effect did not affect analyte quantification in solid matrices in most cases, and the validated method was successfully applied to analyses of short-and long-chain PFAs in various solid matrices.
What problem does this paper attempt to address?