Application of spectrum correction technique in the detection of skin advanced glycation end products

Yong LIU,Yi-kun WANG,Long ZHANG,Ling ZHU,Liang-yuan XU,Ming-xing CHEN,An WANG
DOI: https://doi.org/10.13741/j.cnki.11-1879/o4.2010.02.012
2010-01-01
Guangxue Jishu/Optical Technique
Abstract:Advanced glycation end products are highly correlated with diabetes mellitus. In order to detect skin advanced glycation end products noninvasively, the test is demonstrated by using fluorescence spectrum measurement. The method for correcting fluorescence spectrum is discussed to reject spectrum distortion caused by absorption and scattering from skin. First, irradiating the skin of human arm using light with wavelength of 370 nm, auto-fluorescence is collected by optical fiber detector. Light scanning from 420 nm to 550 nm on the same position, the reflectance curve is acquired. Then intrinsic spectrum of advanced glycation end products is achieved after correcting auto-fluorescence using corresponding reflectance curve. Finally, the intrinsic spectrum of diabetic patients and healthy subjects are analyzed and discussed. Experiment result indicates that the correction is appropriate and the corrected fluorescence spectrum can be used to distinguish diabetic patients and healthy subjects.