Multi-parameter adjustable high-voltage nanosecond pulse generator for inducing apoptosis of tumor cells

Chenguo Yao,Dongyang Zhao,Jianfei Wang,Jian Wang,Ximing Zhang
DOI: https://doi.org/10.19650/j.cnki.cjsi.2010.05.025
2010-01-01
Abstract:In order to study the window parameter selection rules, which could be used to induce apoptosis in tumor cells with an applied nanosecond pulsed electric fields, this paper develops an iatrical, multi-parameter adjustable, portable and intellectualized high-voltage nanosecond square wave pulse generator. The pulse generator adopts pulsed power technology and Blumlein LC pulse forming line network with concentration parameters. The generator is mainly composed of a high voltage DC power supply, a nanosecond pulse forming system and a pulse shaping and counting system. The output pulse amplitude is 4 kV to 15 kV and is continuously adjustable, the pulse width is 50 ns to 500 ns, and the minimum rising time is 10 ns. The repetition frequency is adjustable, the treatment time window is controllable, and the generator has a liquid crystal display and protection function. In order to get the appropriate pulse parameters that can be used to induce apoptosis effectively, human ovarian cancer cell line SKOV3 is used as a research object and the pulse parameters are selected using orthogonal design method. Medical experiments show that the generator is compact, lightweight and has stable performance, which effectively meets the requirement of the parameter selection that could induce apoptosis of tumor cells.
What problem does this paper attempt to address?