A New Method for High-Energy Pulsed Gamma Measurement Within Intense Background X-Rays

Tan Xin-Jian,Ouyang Xiao-Ping,Wang Qun-Shu,Song Zhao-Hui,Kang Ke-Jun,Xia Liang-Bin
DOI: https://doi.org/10.1088/1674-1056/19/1/010703
2010-01-01
Abstract:The accelerator-generating 6.13 MeV pulsed Gamma by 19 F(p, αγ) 16 O reaction usually synchronizes with an intense bremsstrahlung x-ray which has a maximum energy of 1 MeV. This paper proposes a new method, named the scattering and absorbing method, to diagnose the 6.13 MeV Gamma. This method includes two parts: the detector and a scatterer placed in front of the detector. The detector converts the Gamma to electrons and then collects the electrons by a scintillator. In order to restrain the interference of the low-energy background, the scintillator collects the electrons at a small angle. The scintillator is wrapped with electro-absorbing material to absorb the low-energy electrons generated by background x-rays. The theoretical sensitivity ratio of 6.13 MeV Gamma to 1 MeV x-rays is greater than 150. The scatterer is a pretreatment tool to scatter some background x-rays away from the radial beam before they enter the detector. By varying the length, the scatterer can reduce the background x-rays to an acceptable level for the detector.