Technique of image spatial extendibility for cone-beam CT on medical linear accelerator

Jian Zhu,Min Liu,Limin Luo
DOI: https://doi.org/10.3969/j.issn.1001-0505.2010.01.015
2010-01-01
Abstract:The traditional cone-beam CT (CBCT) scanning process on medical linear accelerator was improved. The CBCT image fields were rewrote to extend the image spatial length. The veraci-ties and feasibilities of this technique were validated in the dosimetric methods. Same conformal radi-ation therapy plans were added on planning CT and spatial extended CBCT respectively. Comparing the dose distribution on two different image spaces, difference on relative percentage of total-lung volume receiving dose larger than 20 Gy is 0.5% ±0.22% and on total-lung mean dose is (38.2 ± 8.82) cGy, which mean there are no significant dose distribution difference between planning CT and CBCT. This technique, which enlarges the CBCT field and improves the accuracy of oncology radiation therapy, would not only provide scientific method for adaptive radiation therapy plan evalu-ation but also has good prospect of application and extension. The complications from the additive low-dose radiation to patient's normal tissue and skin should be investigated in future clinical trials.