Development of a Force-Balance Telemetry Device for Total Knee Arthroplasty

DANG Xiao,ZHU Zhong-lin,DING Hui,ZHOU Yi-xin,WANG Guang-zhi
DOI: https://doi.org/10.16156/j.1004-7220.2010.02.012
2010-01-01
Abstract:Objective A new type of tibial-femoral force-balancing telemetry device was developed for measuring force balance on the tibial-femoral contact surface in knee joint. Method The force-balance telemetry device was designed imitating the configuration of tibial part of total knee prosthesis,in which force sensors and telemetry circuit were embedded. During total knee arthroplasty(TKA),this device was implanted as a tibial prosthesis to measure tibial-femoral force while the knee joint was being placed at different angles. The tension balance was adjusted by releasing soft tissue around knee joint. Results The results indicat that this device has advantage of high accuracy for force measurement (r0.98,RMS= 65 g),which is sufficient to meet clinical demand. Conclusions The measurement of force balance on the tibial-femoral contact surface with telemetry device can provide new insight into the estimation of optimal knee prosthesis position and selection of appropriate polyethylene insert.