Experiment on the surface plasma resonance imaging sensor

Feng-ze JIN,Chun-lei DU,Li-fang SHI,Hong-tao GAO,Qi-ling DENG,Jing-lei DU,Yong-kang GUO
DOI: https://doi.org/10.3969/j.issn.1007-2276.2010.02.018
2010-01-01
Abstract:Based on the principle of light intensity modulation sensor of surface plasma resonance (SPR)and the superiority of microscope imaging system with high resolution,a simple structure and low-cost SPR refractive index imaging sensor was fabricated.The main advantages of this new sensor included high sensitivity,free of label and rapid real-time measurement.It Can be widely used for biosensors and biochemical interaction monitors.The reflected light intensity versus refractive indices of different sucrose solutions have been measured experimentally by using the self-made SPR imaging sensor.Several main reasons for affecting the sensitivity of experiment were analyzed which provided a basis for system improvement.Experimental results show that the measurement sensitivity of the new sensor Can reach to 3.6×10-4 RIU.
What problem does this paper attempt to address?