Quantification of somatostatin with a surface plasmon resonance-based biosensor

GAO Jin-Hang,Rui Liu,Huan Tong,Xian Li,Ou Qiang,TANG Cheng-Wei
2010-01-01
Abstract:A quick technique of quantifying somatostatin (SST) with surface plasmon resonance (SPR) on the principle of antigen-antibody-antigen assay was established. The optimal coupling of CM5 sensor chip for SST was indicated as 1231.4 response unit (RU) at pH = 5.0, and the SST concentration of 75 mu g/mL. A negative linear relationship between SST content (20-2000 pg) and RU was obtained using the high molecular mass antibody for SST at 117.7 nmol/L, R = -0.824, p<0.05, coefficient of variation 1.3%. This method can be used to detect SST content at a range from 20 to 2000 pg with good repetition and real-time, which would be useful for clinical diagnosis and intensive care of critically ill patients.