Noninvasive temperature imaging using permanent MRI

Peng Yuhua,Dong Lin,Tang Xiaoying
DOI: https://doi.org/10.1109/ICBBE.2009.5163117
2009-01-01
Abstract:In this paper, temperature mapping by using permanent magnetic resonance imaging was studied. The different temperatures of porcine liver are measured with MRI using Spin-Echo (SE) sequence, which is based on the relationship between temperature difference (ΔT) and average signal amplitude difference (ΔM). The experimental results show that the temperature measured with MRI was basically identical with the digital thermometer measuring. ©2009 IEEE.