Design and Clinical Application of the Interventional Robot THMR-I

Jing Xiong,Ken Chen,Zeyang Xia,Xiangdong Yang,Ping Liang
DOI: https://doi.org/10.1109/aim.2009.5229925
2009-01-01
Abstract:An ultrasound-guided interventional robotic system includes the image navigation system, the interventional robot, and the electro-magnetic tracking unit, among which the interventional robot is a key issue affecting the therapy effect of the interventional therapy. According to the requirements of the clinical interventional therapy, an interventional robot, THMR-I, is designed. The robot combines the SCARA structure and the serial wrist to realize the wide angle active locating and puncture. This paper describes the design aspects of THMR-I concerning the medical functionality, Safety and reliability. The accuracy of the robot system is specifically analyzed theoretically and experimentally. THMR-I has been applied in over ten cyst puncture surgeries, including hepatic cysts and renal cysts, and achieved desired therapy effects, which verified the system's design concepts, reliability, and accuracy.
What problem does this paper attempt to address?