1D-KBA Microscope Using Double-Periodic Multilayer

Baozhong Mu,Zhanshan Wang,Shengzhen Yi,Jingtao Zhu,Xin Wang,Li Jiang,Qiushi Huang,Yuhong Bai
DOI: https://doi.org/10.1117/12.825767
2009-01-01
Abstract:To study the action of shock wave in CH target, one-dimensional grazing incidence KBA microscope for 4.75keV energy was set up. Because of strong absorption in air, 4.75keV energy microscope can just work in vacuum. Accordingly, the alignment and assemblage will be very complicated and difficult. A special multilayer method, using double periodic multilayer, was proposed to solve this problem. This multilayer has high reflectivity not only for 4.75keV x-rays but also for 8keV x-rays at the same grazing incidence angle. It means 1D-KBA microscope has the same light trace for different working energies. Therefore, we can implement the alignment and assembly of 4.75keV system by the help of 8keV x-rays. Because 8keV x-rays is very easy produced by x-ray tube and has strong transmittability in air, the alignment and assemblage process became relatively easy. By now, we have finished the alignment experiment at 8keV and obtained imaging results. The performance is about 2-3 mu m resolution in 250 mu m field of view. It is coincide with the calculation.
What problem does this paper attempt to address?