Nanomagnetic material ferriferrous oxide separation/Enrichment and inductively coupled plasma-atomic emission spectrometry. for determination of arsenic

Shao Xiong Wu,Zhi Xing,CHEN Hong-Bing,Juan Wang,Lu Feng
DOI: https://doi.org/10.3321/j.issn:0253-3820.2009.05.017
IF: 1.193
2009-01-01
Chinese Journal of Analytical Chemistry
Abstract:Using a magnetic separation technique, the novel separation and preconcentration technique for the determination of trace As in water samples was established. Arsenic levels can be determined by ICP-AES. The adsorption behavior of arsenic speciation on nanomagnetic material Fe(3)O(4) was investigated. pH values for solution, time of adsorption, amounts of adsorbent, volumes of solutions and the co-existent ions were considered. At pH 2 - 10, As (III), As (V), monomethylarsenic (MMA) and dimethylarsenic (DMA) were retained on 6.0 mg nanomagnetic Fe(3)O(4) after shaking 2 min. Extractions were performed in test tubes containing 10 g arsenic ions at pH 5, using 6.0 mg nanomagnetic material Fe(3)O(4). They are attracted to a magnetic field and can be separated from solution within a minute with a magnet. The detection limit (3 sigma) was 0.0189 mg/L with relative standard deviations of 1.3%. It was obtained a wide linear range of 0.063 - 20 mg/L and the recovery is 100.3% - 104.8%
What problem does this paper attempt to address?