Abstract:Aiming at noninvasive blood glucose sensing by nearinfrared spectroscopy,the principle of transcutaneous glucose measurement was analyzed.Due to the weak signal and the difficulty in quantifying the physiological noise directly,Monte Carlo simulation was used to compute the light intensity induced by the change in glucose concentration and physiological temperature.And the reliability of simulation computation was validated using glucose aqueous solution.Finally,the influence of actual physiological temperature on the prediction model of glucose concentration was estimated based on the oral glucose tolerance tests of two diabetics.The result showed that,near the normal temperature of 37 ℃,the intensity of diffuse reflectance caused by-0.1 ℃ change in temperature was equivalent to that caused by +2.7 mmol/L change in glucose concentration.Moreover,temperature was one of the most significant error sources for the blood glucose prediction by near-infrared spectroscopy,and the ratio of the error induced by temperature to the total error was more than 50%.