Study of cone beam CT imaging system with Co-60 as the radiation source

Ji Changguo,Duan Chaijie,Wang Haoyu,Jia Pengxiang,Cui Yidong,Bao Shanglian
2008-01-01
Abstract:Image guidance system, helping to target tumor precisely, was successfully used in linear accelerator radiotherapy instrument. Considering the specialty of mechanical structure of gamma-ray stereotactic radiotherapy instrument, we planned to develop a Co-60 based image guidance system. We designed and installed a cone beam CT imaging system with Co-60 as the radiation source. Various CT scanning trajectories (including circle, spiral, saddle, and circle-line) was successfully implemented on this system. The chord-line and VOI based BPF algorithm with a general trajectory and FDK algorithm were used for image reconstruction. We studied the imaging quality of this system using head and chest phantom respectively, and finally use of the Min/Max curvature flow method to eliminate the noise of the reconstructed image. The result showed that the contour and bone boundaries of scanned object can be clearly determined, which supplied the basis of further patient auto-positioning for cobalt-60 gamma-ray stereotactic radiotherapy device.
What problem does this paper attempt to address?