Characteristic analysis of electroporation apparatus for transdermal drug delivery

Ying Lei,Bao Jiali
DOI: https://doi.org/10.3321/j.issn:0254-3087.2008.01.004
2008-01-01
Abstract:The characteristic of an electroporation apparatus for transdermal drug delivery is discussed in the paper. The voltage generator generates DC 400 V from DC 12 V power supply using a DC/DC converter in the apparatus, and charges four storage capacitors 2.2 μF, 22 μF, 47 μF and 100 μF. One of the four capacitors discharges to skin through Ag-AgCl electrodes under the control of a pulse controller, and generates a series of exponential decay pulses on the skin. Four waveforms of the discharge pulses are recorded. The results show that when truncation coefficient is less than 1/5, complete exponential decay pulse can be acquired, and the skin resistance is 3.57 ± 0.66 kΩ. The truncation coefficient is restricted by both transdermal energy and pulse protocols. If parallel resistance and storage capacitance in the system are selected optimally, the highest power supply energy efficiency and the lowest pulse truncation coefficient can be obtained.