Application of total error correction technique in thermal deformation equipment

Li Guihua,Ma Xiushui,Zhang Mei,Fei Yetai
DOI: https://doi.org/10.19650/j.cnki.cjsi.2008.12.016
2008-01-01
Abstract:Error correction technique is a commonly used method to improve experiment measurement precision. But sometimes because the error sources are too many, or the error is not easily measured, the error correction is not complete or can not be used. This paper points out a total error correction method, which takes the whole measurement equipment as one error source to get rid of the measurement error caused by each individual error source, and then the error is measured using an equivalent replacement method and the result is corrected. This method was applied to high-precision thermal deformation measurement equipment, and test result indicates that this method can effectively improve the precision and reliability of the measurement data.
What problem does this paper attempt to address?