Optimum Design of Flat-Field Holographic Concave Grating for Micro-Spectrometers

CHENG Liang,CHEN Yan-ping,ZHU Ruo-bo,YE Zi,YU Fei-hong
DOI: https://doi.org/10.3785/j.issn.1008-973x.2008.02.026
2008-01-01
Abstract:A design method with globe optimization and reverse optimization was presented to achieve a flat-field holographic concave grating used for micro-spectrometer system.Globe optimization algorithm was utilized to solve the nonlinear equation set of flat-field holographic concave grating,which helps the designer freely adjust the weights of specific aberrations during the design process and shorten the design time.Then,the idea of reverse optimization was raised to make little adjustment to the working parameters of intermediate results,which improves the optical performance of the grating.Base on this reverse optimization thought,the possibility of the gratings' multi-working configurations was enunciated.Genetic algorithm was adopted to realize the presented design thought.Based on a TCD1304AP linear CCD detector,a flat-field holographic concave grating used for micro-spectrometer with working wavelength range of 380~780 nm was designed.The results showed that the actual system's spectral resolution was about 0.8 nm when using 50 microns wide slit.
What problem does this paper attempt to address?