Development and Properties of A New TiNi Thoracic Aorta Stent

Yuling Hu,Weidong Miao,Shangdong Xu,Jinfang Guo
DOI: https://doi.org/10.13373/j.cnki.cjrm.2008.05.013
2008-01-01
Abstract:According to the physiological characteristics and performances of human thoracic aorta, a new thoracic aorta stent of TiNi shape memory alloy was finished by knitting, heat treating at 500°C, sewing and so on, which was designed for minimally invasive interventional therapeutic procedure. The stent was made up of TiNi shape memory alloy wires and artificial blood vessel. The stent was smaller gradually in diameter, and had no girder. The mechanical property of TiNi wires and radial holding force of the stent were measured with electron universal test machine and the recovery temperature was measured by DSC (differential scanning calorimetry), and structural design of the stent had been verified by doctors. The results showed that the stents recovered shape completely at 25.9 ± 3°C and radial holding force of the stent distributed between 6 and 8 N. Doctors would operate more easily due to its reducing diameter structure with no girder. The stent had good flexibility and uniform expandability, and was more suitable to meet the physiological characteristics of human thoracic aorta. The results of clinical application were satisfactory.
What problem does this paper attempt to address?