Analytic Calculation of CT Phantom Projections and Its Application in Beam Hardening Correction

汤少杰,牟轩沁,闫浩,杨莹
DOI: https://doi.org/10.3321/j.issn:0253-987x.2008.06.004
2008-01-01
Abstract:To develop an efficient beam hardening correction for medical diagnostic X-ray CT(computed tomography),an analytic calculation method was proposed to simulate polychromatic X-ray projections.The method combines the energy spectrum distribution of X-ray source,geometrical shapes and material compositions of sub-regions of phantoms.Firstly,CT values are mapped into material compositions under the precondition that only human being tissues are considered.And then,boundaries of sub-regions are determined by the proposed analytic method.Finally,projections are calculated according to the energy spectrum distribution of X-ray source.Compared with the linear integral projecting,the proposed method can not only generate more realistic projections,but also provide a unified projection data for beam hardening correction and CT reconstruction.Experiments of FORBILD head phantom show that the CTvalue inconsistency error of bone correction in bone region is lower than the water correction around 4-5Hu,but higher than the monochromatic reconstruction around 5-6Hu.