A Multi-Position Calibration Algorithm for Inertial Measurement Units

Hongliang Zhang,Yuanxin Wu,Meiping Wu,Xiaoping Hu,Yabing Zha
DOI: https://doi.org/10.2514/6.2008-7437
2008-01-01
Abstract:∗As an inertial sensors assembly, the inertial measurement unit (IMU) must be calibrated before being used. This paper proposes a new IMU multi-position calibration algorithm that takes the Earth’s rotation rate and gravity as inputs, and calculates calibration parameters based on the two facts that: 1) norm of the accelerometer measurement vector is equal to the magnitude of gravity; 2) the dot product of gyro measurement vector and accelerometer measurement vector is equal to minus dot product of the Earth’s rotation rate and gravity. Two theorems about the rank of the involved matrix are given to prove the feasibility of estimation. The algorithm features that no high-precise instruments such as a turntable are in principle necessary. Simulations show feasibility of the proposed calibration algorithm.
What problem does this paper attempt to address?