Determination of Captopril in Human Plasma by RP-HPLC

Mao-yi WANG,Ya-lin DONG,Hai-sheng YOU,Wei-hua DONG,Xin ZHENG
DOI: https://doi.org/10.3321/j.issn:1001-2494.2007.07.014
2007-01-01
Abstract:OBJECTIVE: To establish a reversed phase high performance liquid chromatographic method for the determination of captopril in human plasma and to investigate the pharmacokinetics of captopril in Chinese healthy volunteers. METHODS: The drug was extracted from human plasma with ethyl acetate-benzene (1:1). The plasma concentrations of captopril were separated on a HYPERSIL C 18 column (4.6 mm x 250 mm, 5 μm) by Waters 510 HPLC. The mobile phase was a mixture of acetonitrile-1% acetic acid (45:55). The detection wavelength was 257 nm. RESULTS: A good linearity was obtained over the range of 6.25 to 800 μg·L -1 in human plasma. The limit of detection was 6.25 μg·L -1. The recoverys of method were from 101.00% to 108.88%. The precisions of intra-day and inter-day were in the rangs of 4.40%-7.88% and 2.87%-9.34%, respectively. CONCLUSION: The method is sensitive, accurate, reliable, and can be used in clinical research of captopril.
What problem does this paper attempt to address?