Space Precision Orientation Method for Cell Micro-Injection System

Wenming Xi,Hui Zhong,Aquan Guo,Daoheng Sun
DOI: https://doi.org/10.3321/j.issn:0253-987X.2007.07.017
2007-01-01
Abstract:The laser trigonometry measurement method is improved to obtain the data of height of probe with a relative error less than 0.4%, which are integrated in the servoing equation to control micro-manipulator movement in three dimensional space. When micro-manipulator moves in image space, the modified sum of squared difference method is employed to track probe, thus the error of micro-manipulator movement is eliminated and probe moves along the expected track to realize the cell micro-injection. The experimental results show that this method enables to minish error from (11, 10) pixels to (0, 2) pixels to accurately control the micro-manipulator movement in three dimensional space.
What problem does this paper attempt to address?