Modeling and Simulation for Permanent Column in Wireless Endoscope

ZHANG Xing,FANG Liang,LI Guo-li,DENG Fan-li,LI Jian-ping
DOI: https://doi.org/10.3969/j.issn.1004-731x.2007.03.008
2007-01-01
Abstract:The mathematical model for permanent column is a key problem for the location of the wireless endoscope. The permanent column is usually regarded as magnetic-dipole to calculate space magnetic field, but it has a precondition that field radius is far-forth bigger than permanent column radius, and it can not reflect vicinity field. To solve this question, an equivalent magnetic charge method was proposed for calculating magnetic field, which does not need precondition and can reflect permanent column magnetic field correctly, and it’s simple. Finally, the model was stimulated in MATLAB6.5 and the results indicate that the model is feasible and superiority.
What problem does this paper attempt to address?