Measurement Of 1-Aminocyclopropane-1-Carboxylic Acid Synthase Activity From Soybean Etiolated Hypocotyls By High Performance Liquid Chromatography

Guan Yong-He,Jun Yang,Bao Yong-Ming,Ming Liu,Lijia Àn
IF: 1.193
2007-01-01
Chinese Journal of Analytical Chemistry
Abstract:A high performance liquid chromatography method has been developed for the determination of 1-aminocyclopropane-1-carboxylic acid synthase (ACC synthase). The peak time of standard samples of S-aderosine methionine (SAM), adenine and adenosine are determined as 25, 12 and 16 min respectively, by using HPILC with a C-18 (250 min x 4.6 mm, 5 mu m) column and UV detector set at 260 nm. The gradient elution of methanol and 1% aqueous acetic acid was used as the mobile phase. The flow rate was initially set at 0. 25 mL/min for 7 min and subsequently decreased to 0.15 mL/min for 15min, then inereased to 0. 5 mL/min for all system. ACC synthase was extracted from soybean etiolated hypocotyls through homogenization, centrifugation., salt-out and ultrafiltration. ACC synthase was mixed with equal volume SAM standard and water bathed at 30 degrees C for one hour. The system was then analyzed by HPLC and the outcome peak appeared at the same time as adenine standard. Electrospray ionization mass spectrometry ( ESI-MS) analysis confirmed that this system yielded a protonated molecular ion of adenine at m/z 136.1.