The Heat Transfer Enhancement Characteristics of a Novel Tumor Cryoprobe

QI Shou-liang,ZHANG Peng,WANG Ru-zhu,ZHANG Ai-li,XU Xue-min
DOI: https://doi.org/10.3321/j.issn:1006-2467.2006.08.024
2006-01-01
Abstract:The cryosurgery system was developed by using a vapor-liquid separator.The effect of the driven pressure on the temperature of the probe surface was investigated.According to the heat transfer analysis,the flow boiling pattern inside the probe in 17(°C) brine is considered to be the film boiling.Hence,two kinds of heat transfer enhancement techniques for the flow film boiling,the helical coiled wire and the helical mesh,were proposed,and the heat transfer characteristics were studied primarily.The results show that the precooling time of the probe is shortened significantly with the help of the vapor-liquid separator, and decreases with the driven pressure.With the helical coiled wire and the helical mesh,significant thermal oscillation corresponding the density-wave is observed in the precooling time.The average heat transfer coefficient enhances 1.5 times with the mentioned heat transfer enhancement techniques,and the temperature of the 2.5 mm cryoprobe surface can reach-160(°C) in 17(°C) brine within 10 min.
What problem does this paper attempt to address?