Capsule location detection based on magnetoresistive sensor in GI

He Wenhui,Yan Guozheng,Guo Xudong
DOI: https://doi.org/10.19650/j.cnki.cjsi.2006.10.004
2006-01-01
Abstract:The principle and characteristic of HMC1023 magnetoresistive sensor is introduced. The calibration method of the sensor is designed. The application of the sensor in localization of the noninvasive monitoring capsule in GI (gastrointestinal tract) is studied. For capsule localization, the sensor is fixed inside the capsule, and three coils for stimulating magnetic field are fixed outside the body, so the capsule's coordinate value can be calculated from magnetic theory. Experiment indicates that the proposed method is simple and reliable. The method will probably be applied in practice with continuous researching. However, the magnetoresistive sensor must play an important role in the localization system.