A Novel Chemiluminescence System for Producing and Measuring Peroxynitrite Anions

CHEN Jiwu,HU Bin,SU Yu,QIN Haiyan,CAO Jiaqi
DOI: https://doi.org/10.3969/j.issn.1000-3436.2006.01.003
2006-01-01
Journal of Radiation Research and Radiation Processing
Abstract:The study for generating and scavenging peroxynitrite anions (ONOO?) is an urgent task in free radi-cal biology and medicine. Therefore, a novel chemiluminescence system for producing and measuring ONOO? has been established. The optimal conditions for detecting ONOO? were evaluated. The method can be described as fol-lows. A mixture solution containing 10 mmo l / L hydroxylamine, 0. 5 mo l / L NaOH and 1mmo l / L EDTA was stirred vigorously in aerobic conditions for about 3 h to form ONOO?. Certain amount of MnO2 powder was added to the mixture solution to eliminate H2O2 generated in the reaction. Then 10 0 μL antioxidant and 860μL 0.1mo l / L carbon-ate buffer (pH10) were added into a test tube. By adding 40μL filtrated ONOO? solution into the tube, chemilumines-cence (CL) reaction was initiated. Ten seconds after the solution was uniformly mixed, CL intensity of the reaction solution could be determined in a measurement of 10 seconds. The method was tested with tea polyphenol, ascorbic acid, rutin and baicalin, whish are known to have effects of scavenging ONOO?. The tests show that the system is sensitive, low?cost, credible and easy method for screening ONOO?.
What problem does this paper attempt to address?