Analysis of Bolus Chasing Angiography in Peripheral Vascular Disease

LIU Bo-shan,ZHANG Yu-xiang,WANG Jian,WANG Chao,ZOU Ying-hua
DOI: https://doi.org/10.3969/j.issn.1672-8475.2006.06.006
2006-01-01
Abstract:Objective To compare the techniques of bolus chasing angiography(BCA) and digital subtraction angiography(DSA).Methods Sixty-one patients with symptomatic atherosclerotic peripheral disease were randomly assigned to have their lower limbs examined by BCA or DSA.Time of examination,dose of contrast medium,dose-area product(DAP) were measured.Image quality was assessed from the laser printed images by two professors.Results Examination was performed successfully by the two methods.DSA produced better images of the infrainguinal vessels(59/60 vs 52/62),but at a high DAP([134.67±1.34] Gycm~2 vs [69.55±0.74] Gycm~2).Contrast medium dose was higher with DSA([94.58±1.44] ml vs [46.77±0.94] ml).Times was higher with DSA than BCA([47.17±0.56] s vs [32.55±0.28] s).Statistically significant difference was showed(P0.001).Image quality evaluation of iliac,femoral and popliteal vessels was no statistically difference(P0.05).Conclusion BCA is the most effective way which required more technical skill.We suggest that BCA be used as the first choice and be complemented by DSA while local pathological change is detected.
What problem does this paper attempt to address?