Kinetic assay of trypsin with a wireless magnetoelastic sensor

ShiHui Wu,Qingyun Cai,Craig A. Grimes
DOI: https://doi.org/10.1166/sl.2006.010
2006-01-01
Sensor Letters
Abstract:A wireless magnetoelastic trypsin sensor is fabricated by first coating a magnetoelastic ribbon-like sensor with a pH-sensitive polymer and then upon it a layer of trypsin. In response to an externally applied time-varying magnetic field, the magnetoelastic sensor mechanically vibrates at a characteristic frequency that is inversely dependent upon the mass of the attached film. As the magnetoelastic sensor is magnetostrictive, the mechanical vibrations of the sensor launch magnetic flux that can be detected remotely from the sensor using a pickup coil. The trypsin-catalyzed hydrolysis of N-alpha-benzoyl-L-arginine ethyl ester (BAEE) decreases the ambient pH, resulting in the pH-responsive polymer shrinking, and consequently the resonance frequency of the sensor increasing due to the decreased mass loading. The optimum pH and temperature are investigated. The kinetic parameters of both the immobilized and free trypsin are measured. Immobilization of trypsin resulted in the enzyme activity decreasing by about 50%. The magnetoelastic sensor demonstrates a linear shift in resonance frequency with trypsin concentration between 24 U/ml and 360 U/ml with a trypsin detection limit of 24 U/ml, and can be used to detect the substrate BAEE ranging from 0.4 mM to 1.5 mM.
What problem does this paper attempt to address?