An Octadecyl Immobilized Silica Beads Packed Microcolumn Versus Polytetrafluoroethylene Knotted Reactor for Cd(Oh)(2) Precipitate Collection Coupled to Electrothermal Atomic Absorption Spectrometry for Cadmium Screening

Yang Wang,Wang Jianhua
DOI: https://doi.org/10.3321/j.issn:0253-3820.2006.08.006
2006-01-01
Abstract:Octadecyl Immobilized surface was, proved to be an effective collecting medium for cadmium hydroxide precipitate. A thorough scrutiny comparison of precipitate collection and related analytical performances were made by employing a C-18 microcolumn packed in a lab-on-valve system and a polytetrafluoroethylene knotted reactor (PTFE-KR). Cadmium was isolated as hydroxide precipitate and collected on the surface of the C-18 microcolumn or the inner surface of the PTFE-KR. The precipitate was afterwards quantitatively eluted with 20 mu L nitric acid (1%), and the eluate was subsequently transported into the graphite tube of the electrothermal atomic absorption spectrometry (ETAAS) instrument for atomization. With a sampling volume of 600 mu L, the obtained detection limits (ng/L), enrichment factors, retention efficiencies (%), precisions (%, 0.05 mu g/L, n = 7) and sampling frequencies by using a C-18 microcolumn/PTFE-KR were 1.7/3, 28/14, 93/48, 2. 1/4. 7 and 13/20, respectively. The procedure was validated by analyzing cadmium contents in three certified references materials, river sediment (CRM 320), sea lettuce (CRM 279) and frozen cattle blood (GBW 09140). Agreements between the obtained results and the certified values were achieved.