STUDY ON IODOMETRIC DETERMINATION OF H2O2 MASS FRACTION IN PROPYLENE EPOXIDE

ZHAO Ji-gang,SHEN Ben-xian,XIAO Wei-guo,ZHANG Ce,WANG Lei
DOI: https://doi.org/10.3969/j.issn.1001-8719.2005.04.007
2005-01-01
Abstract:The orthogonal design was employed to research the optimized titration conditions for a simulated solution of propylene epoxides. It is concluded that the amount of sample gave the most significant influence to the titration result, while the amounts of both KI and H2SO4 solutions only provided significant influences. The mass fraction (x) of H2O2 measured by titration in propylene epoxide solution was well consistent with its true value if x was in the range of 0.5%-3.3%, and the minimal amount (m) of the solution being titrated was estimated by m = 0.17/x and the amounts of KI and H2SO4 solutions used in titration were between 3-8 ml and 2-7 ml respectively. It is shown that titration results of the solution with different H2O2 contents were always well consistent with their true value under these conditions. The maximum amounts of KI and H2SO4 solutions used in titration were not more than 8 ml and 7 ml respectively at the relative low mass fraction (<0.5%) of H2O2 in propylene epoxide solution, and the minimum amounts of them are not less than 3 ml and 2 ml respectively at the relative high mass fraction (>3.3%) of H2O2, which is then verified by further experiments.