Prosthetic Visual Acuity in Irregular Phosphene Arrays under Two Down-Sampling Schemes: A Simulation Study

Shanqing Cai,Lin Fu,Hui Zhang,Guangshu Hu,Zuoqing Liang
DOI: https://doi.org/10.1109/iembs.2005.1615656
2005-01-01
Abstract:To investigate the effects of phosphene array irregularity in visual prostheses, a model of phosphenes' positional uncertainty was devised and two different image down-sampling schemes, adapting and non-adapting, were proposed. Based on a simulation system, visual acuity tests were given to four normally sighted subjects under seven degrees of array irregularity and the two down-sampling schemes. With the irregularity index increasing from 0 to 0.6, visual acuity fell over a range of 0.22 logMAR (logarithm of minimum angular resolution) for the adapting scheme and 0.47 logMAR for the non-adapting scheme, both monotonously. Comparison between the two down-sampling schemes was made. For low array irregularity, non-adapting down-sampling afforded higher visual acuity, whereas for higher irregular the adapting approach did. Head movements were observed to play significant roles in the acuity tests.
What problem does this paper attempt to address?