A Real-time QRS Complex Detection Algorithm for Mobile ECG Telemonitoring System

Zhu Lingyun,Wu Baoming,Wang Zhengguo,Li Gang,Cao Changxiu,Zhu Xinjian,Yan Qingguang
DOI: https://doi.org/10.3321/j.issn:0254-3087.2005.06.013
2005-01-01
Abstract:A precise real-time QRS complex detection algorithm based on local minima of second derivatives and cross-zero points of first derivatives of ambulatory electrocardiograph (ECG) signals to meet the need of mobile ECG telemonitoring system is developed. The novel algorithm filters power-line interference and most of muscle noise and is insensitive to baseline draft and noise caused by mobile communication. The QRS complex recognition thresholds, which can revise themselves according to the detected values and vary with the analyzing signals, are designed. For the normal and clinical patient, this algorithm correctly detects up 99.8 percent of the QRS complex of ambulatory ECG signals.
What problem does this paper attempt to address?