Abstract:A non-invasive and continuous blood glucose monitoring would be of great advantage for diabetic patients. Many techniques have been proposed for the purpose. But so far, none of these methods has been proven to be reliable and precise enough for in vivo monitoring.In non-invasive glucose measurement using near-infrared (NIR) spectroscopy, the difficulty is that the spectral variations due to the glucose concentration are extremely small compared with other sources of variations. Therefore extracting the variation signal of glucose in complicated background is challenging.We investigated the relationship between sample complexity and prediction accuracy, which was the fundamental research of non-invasive sensing and a kind of method to determine whether the OGTT or non-invasive sensing can achieve the required accuracy of clinic. A series of in vitro experiments had been conducted with different complex samples and same measurement system to analyze the relation between the sample complexity and the prediction accuracy, and some conclusions had been drawn. In general, the increase of sample complexity doesn't lead to the distinct increase of prediction error.
What problem does this paper attempt to address?