Fluorescent Reaction Of Ascorbic Acid And Cerium(Iv) And Its Analytical Application

Changlun Tong,Guanghong Xiang,Weiping Liu
DOI: https://doi.org/10.3321/j.issn:0253-3820.2005.01.021
IF: 1.193
2005-01-01
Chinese Journal of Analytical Chemistry
Abstract:Experiments indicated that cerium 0 ion which could not emit fluorescence was reduced to cerium(M) ion which could emit characteristic fluorescence by ascorbic acid in water solution, while sodium triphosphate was added the fluorescence intensity of the system could be greatly enhanced. From this, an indirect sensitive method for determining ascorbic acid was developed. The fluorescence intensity of the system was measured in a 1 cm quartz cell with excitation and emission wavelengths of 303 nm and 353 nm, respectively. The results showed that the fluorescence intensity of the system presented a linear relationship with the concentration of ascorbic acid in the range of 1.0 X 10(-7) similar to6.0 X 10(-6) mol/L, the correlation coefficient was 0.9989, the detection limit (S/N = 3) was 1.0 x 10(-8) mol/L. The presented method was used to determine ascorbic acid sample and vitamin C tablet, the results were satisfactory.
What problem does this paper attempt to address?