Anodic Stripping Voltammetric Immunoassay Based on Silver Enhanced Gold Label for Determination of Complement Iii

X Chu,X Fu,GL Shen,RQ Yu
DOI: https://doi.org/10.3321/j.issn:0251-0790.2005.09.042
IF: 3.1
2005-01-01
Chemical Research in Chinese Universities
Abstract:A novel, sensitive anodic stripping volammetric immurroassay was developed based on silver-enhanced immuno-gold label. The immunoreaction was performed in a polystyrene microwell by using the sandwich formal. Primary antibodies specific for C-3 (complement III) were adsorbed passively on the walls of a polystyrene microwell. The C-3 analyte was first captured by the primary antibody and then sandwiched by a secondary colloidal gold-labeled antibody. The addition of the silver enhancement solution results in the precipitation of a large amount of silver on colloidal gold labels due to the catalytic reduction which, after silver metal dissolution in an acidic solution, was determined by anodic stripping voltammetry(ASV) at a glassy-carbon electrode. The influence of some immunoassay conditions upon the anodic stripping peak current was examined and optimized. The anodic stripping peak current depended linearly on the logarithm of C, mass concentration over the range of 7.2 ng/mL to 7.33 mu g/mL and a detection limit as low as 7 ng/mL is achieved. The anodic stripping volammetric immunoassay was applied to the determination of C, concentration in human serum with satisfactory results.
What problem does this paper attempt to address?