Study on in vitro pyrogen test using human whole blood

黄清泉,刘文英,丁黎,奚廷斐,李冠民
DOI: https://doi.org/10.3321/j.issn:1001-2494.2004.05.020
2004-01-01
Abstract:OBJECTIVE: To optimize the in vitro pyrogen test using human whole blood. METHODS: The levels of cytokines released in blood (namely endogenous pyrogens) were detected by double antibody enzyme-linked immunoassay. RESULT: TNF-α, IL-β and IL-6 were selected as the indicators. The optimal conditions were established as following respectively: reaction time 6 h; Blood dilution times: two; preservation temperature of blood at 4 °C and preservation time 4 h; ratios of blood to test solution in volume. 4:1. CONCLUSION: The results provided basis for the standardization of the in vitro pyrogen test using human whole blood.
What problem does this paper attempt to address?