STUDY OF THE TRANSVALVULAR GRADIENT OF A NEWLY CHINA-MADE BILEAFLET HEART VALVE PROSTHESIS

徐嗣卫,章志量,陈如坤,董爱强
DOI: https://doi.org/10.3969/j.issn.0258-8021.2004.06.012
2004-01-01
Abstract:To assess the transvalvular gradient of a newly China-made bileaflet heart valve prosthesis, the hydrodynamics were performed in the test and the hemodynamic performance were measured in the animals and patients. Methods 1. Test: the prosthesis has been tested in a pulsatile flow simulator in the aortic position. 2. Animal experiments: Six sheep that underwent mitral replacement with the prosthesis of 21mm were measured by open cardiac catheterization intraoperatively. Doppler echocardiography and open cardiac catheterization under dobutamine stress was performed in two sheep 2.5 years post-implant. 3. Patients measurements: 14 cases of aortic valve and 10 cases of mitral valve in the patients who underwent valve replacement with the heart valve prosthesis have been measured by open cardiac catheterization and Doppler echocardiography. Result: 1. Test: The newly heart valve prosthesis had the lower mean transvalvular gradient which was in the range of below 10 mmHg at any given tissue annulus diameter. 2.Animal experiments : The transvalvular gradients of 6 sheep were 5.2mmHg±1.7mmHg intraoperatively and 2 sheep 2.5 years post-implant were 6.1mmHg±0.3mmHg by catheterization. 3. Patients measurement: The mean transvalvular gradients in the aortic position measured by means of catheterization and echocardiography were 6.26mmHg-4.10mmHg and 9.42mmHg- 7.48 mmHg,respectively. The gradients in the mitral position were 2.10mmHg-1.9mmHg and 5.281mmHg-4.10mmHg,respectively. Conclusion The results of this study demonstrated that the newly valve prosthesis has excellent hemodynamic performance.
What problem does this paper attempt to address?