[Determination of Epimedin C and Icariin in Herba Epimedii by HPLC].

Ming-quan Wang,Zhi-ming Bi,Ping Li,Hui Ji,Fei-long Cheng
DOI: https://doi.org/10.3321/j.issn:1001-5302.2003.11.009
2003-01-01
Abstract:OBJECTIVE To establish a RP-HPLC method for the determination of epimedin C and icariin in Herba Epimedii. METHOD Chromatographic conditions: Hypersil BDS-C18 column, acetonitrice and water containing 0.05% phosphoric acid as gradient eluents, G1315A photodiode array detector at 272 nm. RESULT The average recovery rate of epimedin C was 99.7%, RSD 1.5% (n = 9), icariin 102.5%, RSD 1.1% (n=9). CONCLUSION The method is reliable, stable and well repeatable. It provides the useful information for quality evaluation.
What problem does this paper attempt to address?