Theoretic and clinical study on refraction alteration induced by intraocular silicone oil

赵明威,蒋宇振,黎晓新,姜燕荣
2003-01-01
Abstract:Objective: To study the refractive changes induced by intraocular silicone oil. Methods: 1 The theoretical study based on five schematic, eyes and four shapes of silicone oil bubble models in aphakic eyes was taken to evaluate the relation between ocular refraction and the volume of silicone oil. 2 Refraction was performed on 34 eyes before and after silicone oil extraction. Results: 1 Theoretically, 5 types of phakic schematic eyes showed 6.07-8.39 D increase in spectacle refraction induced by intraocular silicone oil. There was no refractive changes in aphakic schematic eye with flat aqueous fluid/silicone oil interface, but a -5.47 D alteration in spectacle refraction when silicone oil formed a same radius of globe as vitreous cavity and refraction changed with the volume of silicone oil in excessive or relatively inadequate amount of silicone oil injection eyes. 2 Clinically, spectacle refraction induced by intraocular silicone oil increased by 6.905 ± 1.403 D in phakic eyes, decreased by -6.58 ± -3.38 D in aphakic eyes and -9.10 ± -4.42 D in aphakic buckled eyes. Conclusions: 1 Intraocular silicone oil could induce hyperopia in phakic eyes and myopia in aphakic eyes. 2 Helmholtz Schematic Eye proves to be the most suitable model to study the refractive effect of intraocular silicone oil in phakic eyes. (3) Refractive effect of intraocular silicone oil in aphakic eyes is closely related with the volume of silicone oil.