Separation Investigation Of Antisense Phosphorothioate Oligonucleotides Using Capillary Electrophoresis With Linear Polyacrylamide

Bin Deng,Guoan Luo,Jun Wang,Xiangyun Wu
DOI: https://doi.org/10.3321/j.issn:0253-3820.2003.03.024
IF: 1.193
2003-01-01
Chinese Journal of Analytical Chemistry
Abstract:The capillary electrophoresis (CE) with linear polyacrylamide (LPA) was employed to separate the phosphorothioate oligonucleotides (PS-ODNs). The content of 10% linear polyacrylamide was used as a replaceable sieving matrix and 100 mmol/L Tris-borate buffer containing 7 mol/L urea at pH 8.2 was chosen as running buffer. The running voltage was designed at 330 V/cm and injection time was controlled at 300 V/cm for 5 s. During all these procedures, electronic polar of CE would be reversed and LPA freshly replaced after using for every four times. P100-a 20-mer antisense PS-ODNs, contained primarily ( n - 1), ( n - 2), ( n + 1), and ( n + 2) deletion or appending sequences, had been separated with single-base resolution when using this polymer system. The separations of target PS-ODNs of P100 and P101, which was designed for anti-lung cancer, were investigated and their separation conditions were discussed here. The relative standard deviations of repeatability ( n = 6) between migration time and peak area were 0.4% and 0.5%, 0.5% and 0.5% for P100 and P101, respectively. Almost the same results would be observed for the relative standard deviations of intra-day and inter-day precision as well. These results showed that the capillary eleetrophoresis of LPA was very suitable for the separation and analysis of the PS-ODNs.