Evaluation of the Effects of the Ph Value of Buffer on Capillary Electrophoresis-Electrospray Ionization-Mass Spectrometry Signal Intensity

梁振,段继诚,张维冰,张玉奎
DOI: https://doi.org/10.3321/j.issn:1000-8713.2003.01.003
2003-01-01
Abstract:In recent years, the hyphenation of capillary electrophoresis (CE) and electrospray ionization mass spectrometry (ESIMS) has been widely applied to some research fields with high efficiency and high accuracy. The effects of pH of the mixed solution consisting the running buffer and the sheath flow on the ESIMS signal intensity were investigated experimentally and theoretically with a mixture of two model peptides, histidineprolineleucine (HisProLeu) and alaninetryptophan (AlaTrp). The running buffer and the sheath flow were found to affect the signal intensity in a synergistic fashion. The ESIMS signal intensity of HisProLeu increased with the H+ concentration of the mixed solution ranging from 0.5 to 2.3 mmol/L, but decreased with the H+ concentration range of 2.3 to 3.09 mmol/L. It is concluded that the ESIMS signal intensity of these peptides increases at first and then decreases subsequently with the increase of the concentration of H+ of the mixed solution, and therefore, there exist an optimal pH value of the mixed solution to maximize the MS response.